Description

Small-sized measuring converters with compensation of magnetic fluxes of DSEN-MI DSET-MI, DSER-MI, are intended for measurement of excessive pressure, razvyazheniya/to a difference of pressure, level and an expense and continuous transformation of the measured parameter to the unified current output signal. Converters can be used in a koshlekta .so by secondary devices and other devices of automatic equipment and control in control systems working from a standard entrance signal 0-5, 0-20 or 4-20 ma.

On resistance to climatic influences have executions:

In, category of placement 4.1 in accordance with GOST 15150-69, but for work at a temperature from 15 to 35 °C (the main option);

UHD or 0, categories of placement 4.2 and execution of T, category of placement 2.3 in accordance with GOST 15150-69, but for work from plus 5 to plus 60 °C.

Description

The principle of action the converter"' consists in * following: The measured parameter will be transformed by a sensitive element to proportional movement of a permanent magnet which creates the operating influence in the form of a magnetic flux and causes change of magnetization of cores.

At the same time there is a mismatch signal which operates an output signal of the amplifier. Usilenna ** the signal comes to the line of remote transfer and at the same time to a winding of feedback which creates the magnetic flux compensating influence of the operating magnetic flux.

Converters consist of the following knots: the measuring block, the magnetomodulation converter and the intensifying device which consists of a payment and the power supply unit.

Structurally converters differ in a design of a sensitive element of the measuring block where serve as a sensitive element in the DME-MI, DMEU-MI, DMER-M, DMER-MI converters - a membrane; in the DSE-MI, DSEYI, DSET-MY, DSER-M, DSER-MI converters - the bellow valve; in the MPE-MI converters - a spring.

Main technical characteristics

Limit of the allowed main error of converters, %-0,6-1,0-1,5.

Note: Converters with a limit blundered - 0,6% are delivered only in reasonable cases in coordination with manufacturer if the volume of their release does not exceed 10% of the general release of converters since they demand selective assembly of knots and details and long technological extra earnings.

Extremely allowed working excessive pressure, MPa: difmanometra. DSE-MI, DSER-M, DSER-MI - 0,1 difmanometra of DME-MI, DMER-M, DMER-MI, DME-MI-K, DMER-MY-K, DMEU-MI - 2,5; Yu; 16; 25; 32; 40

Top limits of measurement of excessive pressure: DSEN-MI difmanometra - 1,0; 1,6; 2,5; 4,0 kPa MPE-MI, MPE-MI K manometers - 100, 160, 250, 400, 600 kPa; 1,0; 1,6; 2,5; 4,0; 6,0; 10; 16; 25; 40; 60 MPas.