Description
Differential manometers (difmanometr) bellows-sealed are intended for measurement:
a) an expense of liquid and gaseous environments by a method of variable difference of pressure (flowmeters);
b) differences of pressure of liquid and gaseous environments (perepadomer);
c) level of the liquid environments which are under atmospheric, vakuummetrichesky or excessive pressure (level gages).
DSS-711 - difmanometr recording with the drive of a chart disk from the electric motor;
DSS-712 - difmanometr recording with the drive of a chart disk from the hour mechanism;
Technical characteristics of the differential manometer
- Extremely allowed working excessive pressures, kgf/cm2: 63; 160; 250; 320
- Top limits of measurements of manometrical part of difmanometr
DSS-711, DSS-712, kgf/cm2
6; 10; 16; 25; 40; 60; 100; 160 - on excessive pressure to 63 and 160 kgfs/cm2
16, 25, 40, 60, 100, 160, 250, 400 - on excessive pressure to 250 and 320 kgfs/cm2 - Limit nominal differences of pressure, kgf/cm2
0,063; 0,1; 0,16; 0,25; 0,4; 0,63; 1; 1,6; 2,5 - on excessive pressure to 63 and 160 kgfs/cm2
0,4; 0,63; 1,6; 2,5; 4; 6,3-on excessive pressure to 250 and 320 kgfs/cm2 - Classes of accuracy 1; 1,5
- Top limits of measurements of the differential manometer:
a) difmanometrov-flowmeters a*10n where and - one of row numbers, n-whole (positive or negative) number or zero get out of a row A =
b) difmanometrov-perepadomer have to correspond to limit nominal pressure differences 1; 1,25; 1,6; 2; 2,5; 3,2; 4; 5; 6,3; 8
c) difmanometrov-level gages get out of a row,
63; 100; 160; 250; 400; 630; 1000; 1600; 2500; 4000; 6300; 10000; 16000 cm
0,63; 1; 6; 2,5; 4; 6,3; 10; 16; 25; 40; 63; 100; 160 m
d) DSP-US difmanometrov-level gages get out of a row
160; 250; 400; 630; 1000; 1600 cm - on excessive pressure of 25 kgfs/cm2 - Time of one turn of the chart, p 24
- Power supply of difmanometr of DSS-711 from the alternating current main:
- tension, In
- frequency, Hz 50 ± 1 - Ambient temperature, °C
DSS-711 - 10... +50
DSS-712 - 10... +50;-30... +50 - Relative humidity, % to 80
- Are made on TU 25-7310.0063-87
- OKP code 42 1253
- Overall dimensions, mm:
difmanometr recording 280 x 340 x 230
showing 195x153x136
showing signaling 235x205x165 - Weight, kg, no more than 16